BLOOD PRESSURE MEASUREMENT SYSTEM BASED ON OSCILLOMETRIC METHOD

Jessica Bolaños Olvera, Roque A. Osornio Rios, Rosalia Reynoso Camacho

Resumen


Abstract

This work presents a methodology for the non-invasive measurement of blood pressure, the algorithm for the obtention of blood pressure (systolic, mean and diastolic pressures) is based on the oscillometric method. In oscillometric method the pressures are determined by applying a mathematical criterion to the oscillometric index pulse. In this work the slope criteria to calculate systolic and diastolic pressures is used. The developed system is capable of obtaining the blood pressure parameters for humans and rats, where blood pressure measurements in Wistar rats are realized as part of the development of other scientific research in Chemistry field.

Keywords: Blood pressure, diastolic pressure, non-invasive, oscillometric method, systolic pressure.


SISTEMA DE MEDICIÓN DE LA PRESIÓN ARTERIAL BASADO EN EL MÉTODO OSCILOMÉTRICO


Resumen

Este trabajo presenta una metodología para la medición no invasiva de la presión arterial, el algoritmo para la obtención de la presión arterial (presión sistólica, media y diastólica) es basado en el método oscilométrico. En el método oscilométrico las presiones son determinadas aplicando un criterio matemático al índice de pulso oscilométrico. En este trabajo es usado el criterio de pendientes para calcular la presión arterial sistólica y presión arterial diastólica. El sistema desarrollado es capaz de obtener los parámetros de presión arterial tanto para humanos como para ratas tipo Wistar, donde las mediciones de la presión arterial en las ratas Wistar se realizan como parte del desarrollo de otras investigaciones científicas en el campo de Química.

Palabras claves: Método oscilométrico, no invasivo, presión arterial, presión diastólica, presion sistólica


Texto completo:

248-265 PDF

Referencias


Aha Committee, Recommendations for standardizations of leads and specifications for instruments in ECG and VCG. Circulation, 52, 11-251975, 1975.

Ahmad, S., Bolic, M., Dajani, H., Groza, V., Batkin, I., & Rajan, S., Measurement of heart rate variability using an oscillometric blood pressure monitor. IEEE transactions on instrumentation and measurement, 59(10), 2575-2590, 2010.

Ball-Llovera, A., Del Rey, R., Ruso, R., Ramos, J., Batista, O., & Niubo, I., An experience in implementing the oscillometric algorithm for the noninvasive determination of human blood pressure. In Engineering in Medicine and Biology Society, 2003. Proceedings of the 25th Annual International Conference of the IEEE (Vol. 4, pp. 3173-3175). IEEE.September, 2003.

Broten, T. P., Kivlighn, S. D., Harvey, C. M., Scott, A. L., Schorn, T. W., & Siegl, P. K. S., Techniques for the measurement of arterial blood pressure. Measurement of cardiovascular function, 1997.

C. A. Ramírez, Algoritmo para el cálculo de la presion sistolica y diastolica en el ventrículo izquierdo. Presented in Memorias II Congreso Latinoamericano de ingeniería biomédica, Habana, Cuba, May. 2001.

Cong, P., Chaimanonart, N., Ko, W. H., & Young, D. J., A wireless and batteryless 10-bit implantable blood pressure sensing microsystem with adaptive RF powering for real-time laboratory mice monitoring. IEEE Journal of Solid-State Circuits, 44(12), pp. 3631-3644, 2009.

Cuesta, Medición de la Tensión Arterial, Dept. Enfermería, Valencia Univ., España, Sep. 2004.

Escobar-Restrepo, B. Y., Sistema para la medición de la presión arterial continua no invasiva sin brazalete, Doctoral dissertation, Biomédica, Mecatrónica y Mecánica, 2014.

Feng, M., & DiPetrillo, K., Non-invasive blood pressure measurement in mice. Cardiovascular Genomics: Methods and Protocols, pp. 45-55, 2009.

Fernández-Abascal, E. G., El tiempo de tránsito del pulso: un índice de cambios en la presión arterial. Estudios de Psicología, 6(21), pp. 21-33, 1985.

Gamboa, W., Rodriguez, L., Cháves, A., de Colombia, F. C., & de Bioingeniería, G., Dispositivo Digital para el registro continuo de presión arterial de forma no invasiva y ambulatoria. In VII Congreso de la Sociedad Cubana de Bioingeniería, 2007.

González, Significación de los ruidos de la presion sanguínea. Sociedad Mexicana para el estudio de la hipertensión arterial, 2008.

Hernández, El modelo en las investigaciones biomédicas. Biomedicina, vol. 2, no. 3, pp. pp. 252-256, 2006.

Infante-Vázquez, O., Sánchez-Torres, G., Martínez-Memíje, R., Flores-Chávez, P., Pastelin-Hernández, G., & Sánchez-Miranda, M., Medición de la presión arterial utilizando el retardo en el pulso distal, Rev Bras Eng Bioméd, 13, pp. 81-92, 1997.

Kramer, K., Voss, H. P., Grimbergen, J. A., Mills, P. A., Huetteman, D., Zwiers, L., & Brockway, B., Telemetric monitoring of blood pressure in freely moving mice: a preliminary study, Laboratory Animals, 34(3), pp. 272-280, 2000.

Leong, X. F., Ng, C. Y., & Jaarin, K., Animal models in cardiovascular research: hypertension and atherosclerosis. BioMed research international, 2015.

Malkoff, J., Non-invasive blood pressure for mice and rats. Animal Lab News, Kent Scientific Corporation, pp. 1-12, 2005.

Mexicana, N. O. NOM-062-ZOO-1999, Especificaciones Técnicas para la producción, cuidado y uso de los animales de Laboratorio. México: Diario Oficial de la Federación, 1999.

Mills, P. A., Huetteman, D. A., Brockway, B. P., Zwiers, L. M., Gelsema, A. M., Schwartz, R. S., & Kramer, K., A new method for measurement of blood pressure, heart rate, and activity in the mouse by radiotelemetry. Journal of Applied Physiology, 88(5), pp. 1537-1544, 2000.






URL de la licencia: https://creativecommons.org/licenses/by/3.0/deed.es

Barra de separación

Licencia Creative Commons    Pistas Educativas está bajo la Licencia Creative Commons Atribución 3.0 No portada.    

TECNOLÓGICO NACIONAL DE MÉXICO / INSTITUTO TECNOLÓGICO DE CELAYA

Antonio García Cubas Pte #600 esq. Av. Tecnológico, Celaya, Gto. México

Tel. 461 61 17575 Ext 5450 y 5146

pistaseducativas@itcelaya.edu.mx

http://pistaseducativas.celaya.tecnm.mx/index.php/pistas